FDA Adverse Event Injury Summary report: N

NEEDLE ELECTRODE MONO Ø 5MM WL 340MM

MDR report key: 18793565 · Received February 27, 2024

Report

Report Number
9611102-2024-00007
Event Type
Injury
Date Received
February 27, 2024
Date of Event
January 25, 2024
Report Date
February 27, 2024
Manufacturer
RICHARD WOLF GMBH
Product Code
GEI
PMA / PMN Number
K964329
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION HAS SHOWN THAT THE AFFECTED DEVICE, TYPE 8383.426, WAS SUBJECTED TO A THIRD -PARTY REPAIR BY MEDWORX GMBH. IN THIS THIRD-PARTY REPAIR, A DIFFERENT INSULATION MATERIAL AND A DIFFERENT PLUG WERE USED. THESE FACTS DO NOT CORRESPOND TO THE DESIGN OF THE NEEDLE ELECTRODE MONO 0 5MM NL 340MM, TYPE 8383.426 FROM RICHARD WOLF GMBH. THEREFORE WE, RICHARD WOLF GMBH, CANNOT MAKE ANY FURTHER EVALUATIONS OR OTHER STATEMENTS. MEDWORX GMBH IS NOT AUTHORIZED BY RICHARD WOLF (RW) TO CARRY OUT REPAIRS OR SERVICE ON RW PRODUCTS. THIS WAS AN UNAUTHORIZED REPAIR. THE PRODUCT NO LONGER COMPLIES WITH THE RW SPECIFICATIONS AND IS THEREFORE TO BE TREATED AS A THIRD-PARTY PRODUCT WITHIN THE SCOPE OF THE EU MDR (MDCG2O23-3).

Description of Event or Problem · 0

ON 29/JAN/2024, RICHARD WOLF GMBH (RWGMBH) WERE INFORMED BY THE FEDERAL INSTITUTE FOR DRUGS AND MEDICAL DEVICES (BUNDESINSTITUT FÜR ARZNEIMITTEL UND MEDIZINPRODUKTE, BFARM) ABOUT A USER REPORT FROM THE "KLINIKUM ST. MARIEN AMBERG". BFARM CASE NUMBER: (B)(4). THE ISSUE IS IN REGARDS WITH A NEEDLE ELECTRODE MONO Ø 5MM WL 340MM, PART ID: 8383.426, BATCH# 1468475. ACCORDING TO THE REPORT, WHEN USING THE MONOPOLAR HOOK, THE INSULATION FELL OFF DURING "STROBING". PARTS OF THE INSULATION COULD BE REMOVED FROM THE ABDOMINAL CAVITY, BUT IT IS NOT 100% CERTAIN THAT ALL OF IT WAS ACTUALLY REMOVED. THERE WAS NO INJURY FOR PATIENT OR OTHER PERSONAL. THE TREATMENT COULD BE COMPLETED WITHOUT CRITICAL TIME DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
444838 NEEDLE ELECTRODE MONO Ø 5MM WL 340MM GEI RICHARD WOLF GMBH 8383426 1468475

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other