FDA Adverse Event
Malfunction
Summary report: N
DELTA
MDR report key: 1879299
·
Received October 19, 2010
Report
- Report Number
- 1220063-2010-00059
- Event Type
- Malfunction
- Date Received
- October 19, 2010
- Report Date
- September 21, 2010
- Manufacturer
- GRAGER MEDICAL SYSMTES, INC. (IT/M)
- Product Code
- MHX
- PMA / PMN Number
- UNK
- Removal / Correction Number
- Z-2283/4-2010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
AN INVESTIGATION WAS INITIATED WITH THE SUPPLIER OF THE SWITCH PANEL. IT WAS DETERMINED THAT THIS CONDITION IS DUE TO A CONTAMINATION PRESENT IN THE FIXED KEY PANEL MATERIAL. A RISK ANALYSIS WAS PERFORMED. DRAEGER MEDICAL INITIATED A RECALL ON (B)(6) 2010 TO ADDRESS THIS PROBLEM. NO PT DEATHS OR INJURIES HAVE BEEN REPORTED AS A RESULT OF THIS CONDITION.
Description of Event or Problem · 1
REFERENCE: MDR 1220063-2010-00054. IT WAS REPORTED THAT THE BEZEL IS CAUSING THE MONITOR TO RANDOMLY PRINT AND SHUT OFF. THERE WAS NO PT INJURY REPORTED. DRAEGER REFERENCE NUMBER: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DELTA | PT MONITOR | MHX | GRAGER MEDICAL SYSMTES, INC. (IT/M) | UNK | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | NO |