FDA Adverse Event Malfunction Summary report: N

DELTA

MDR report key: 1879299 · Received October 19, 2010

Report

Report Number
1220063-2010-00059
Event Type
Malfunction
Date Received
October 19, 2010
Report Date
September 21, 2010
Manufacturer
GRAGER MEDICAL SYSMTES, INC. (IT/M)
Product Code
MHX
PMA / PMN Number
UNK
Removal / Correction Number
Z-2283/4-2010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

AN INVESTIGATION WAS INITIATED WITH THE SUPPLIER OF THE SWITCH PANEL. IT WAS DETERMINED THAT THIS CONDITION IS DUE TO A CONTAMINATION PRESENT IN THE FIXED KEY PANEL MATERIAL. A RISK ANALYSIS WAS PERFORMED. DRAEGER MEDICAL INITIATED A RECALL ON (B)(6) 2010 TO ADDRESS THIS PROBLEM. NO PT DEATHS OR INJURIES HAVE BEEN REPORTED AS A RESULT OF THIS CONDITION.

Description of Event or Problem · 1

REFERENCE: MDR 1220063-2010-00054. IT WAS REPORTED THAT THE BEZEL IS CAUSING THE MONITOR TO RANDOMLY PRINT AND SHUT OFF. THERE WAS NO PT INJURY REPORTED. DRAEGER REFERENCE NUMBER: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DELTA PT MONITOR MHX GRAGER MEDICAL SYSMTES, INC. (IT/M) UNK NA

Patients

Seq Age Sex Outcome Treatment
1 UNK NO