FDA Adverse Event Injury Summary report: N

CERTAS PLUS INLINE VALVE ONLY

MDR report key: 18787811 · Received February 27, 2024

Report

Report Number
3013886523-2024-00051
Event Type
Injury
Date Received
February 27, 2024
Date of Event
January 29, 2024
Report Date
April 11, 2024
Manufacturer
INTEGRA LIFESCIENCES SWITZERLAND SAR
Product Code
JXG
UDI-DI
10886704079002
PMA / PMN Number
K143111
Report Source
Manufacturer report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE CERTAS VALVE (ID (B)(6)) WAS RETURNED FOR EVALUATION. DEVICE HISTORY RECORD (DHR) - THE PRODUCT CODE (B)(6) WITH LOT 200113, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK. FAILURE ANALYSIS - THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS AT SETTING 3. THE CATHETERS WERE IRRIGATED, NO OCCLUSION NOTED. THE VALVE WAS VISUALLY INSPECTED, NEEDLE HOLE WAS NOTED IN THE NEEDLE CHAMBER. THE VALVE WAS HYDRATED. THE VALVE WAS LEAK TESTED AND ONLY LEAKED FROM THE NEEDLE HOLE IN THE NEEDLE CHAMBER. THE VALVE PASSED THE TEST FOR PROGRAMMING, OCCLUSION, SIPHON GUARD AND PRESSURE. ROOT CAUSE ANALYSIS - NO ROOT CAUSE COULD BE DETERMINED AS THE TECHNICIAN COULD NOT CONFIRM ANY PROBLEM WITH THE VALVE AT THE TIME OF INVESTIGATION.

Additional Manufacturer Narrative · 0

AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

N/A.

Description of Event or Problem · 0

A FACILITY REPORTED A CERTAS VALVE (ID 828800PL) HAD BEEN PROGRAMMED INCORRECTLY WITH A MEDTRONIC SHUNT BY ACCIDENT AND THE PHYSICIAN SAID IT WAS FAULTY. THE VALVE WAS IMPLANTED IN A 69 YEAR OLD MALE PATIENT ON (B)(6) 2018 WITH SETTING OF 3. THE VALVE FAILURE WAS DISCOVERED WHEN PATIENT WAS ADMITTED IN THE EMERGENCY ROOM DUE TO DIZZINESS. THE VALVE WAS EXPLANTED AND REPLACED WITH A NEW VALVE ON (B)(6) 2024. PATIENT IS DOING GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
119114 CERTAS PLUS INLINE VALVE ONLY CERTAS PLUS JXG INTEGRA LIFESCIENCES SWITZERLAND SAR 200113 10886704079002

Patients

Seq Age Sex Outcome Treatment
1 69 YR Male