FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 18777806 · Received February 26, 2024

Report

Report Number
2245578-2024-00049
Event Type
Malfunction
Date Received
February 26, 2024
Date of Event
February 14, 2024
Report Date
April 16, 2024
Manufacturer
ABBOTT POINT OF CARE
Product Code
JGS
UDI-DI
10054749001917
PMA / PMN Number
K053110NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON (B)(6) 2024. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AN, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE, FOR SUPPRESSED RESULTS AND FOR POINTS OUTSIDE TOTAL ALLOWABLE ERROR (EA) FOR ALL ANALYTES EXCEPT TOTAL CARBON DIOXIDE (TCO2) IN WHOLE BLOOD (WB). A DEFICIENCY HAS BEEN IDENTIFIED FOR THIS ISSUE WHICH WILL BE INVESTIGATED IN QUALITY RECORD (QR) (B)(4).

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 0

ON 15-FEB-2024, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED DISCREPANT CREA RESULTS ON AN 82 YEAR OLD FEMALE PATIENT WITH ABDOMINAL PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD; DATE; COLLECTED; TESTED RESULT (UNITS) & SAMPLE. I-STAT, (B)(6) 2024, 1320, 1323, 360 UMOL/L & A. PICCOLO, (B)(6) 2024, 1320, 1323, 62 UMOL/L & A. I-STAT, (B)(6) 2024, 1350, 1354, 346 UMOL/L & B. PICCOL0, (B)(6) 2024, 1350, 1354, 68 UMOL/L & B. LAB, (B)(6) 2024, 1350, 1536, 74 UMOL/L & B. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
298553 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE JGS ABBOTT POINT OF CARE NA H23340 10054749001917

Patients

Seq Age Sex Outcome Treatment
1 82 YR Female