4D INTEGRATED TREATMENT CONSOLE
Report
- Report Number
- 2916710-2010-00118
- Event Type
- Malfunction
- Date Received
- October 13, 2010
- Date of Event
- July 22, 2010
- Report Date
- July 22, 2010
- Manufacturer
- VARIAN MEDICAL SYSTEMS
- Product Code
- IYE
- PMA / PMN Number
- K050479
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION HAS IDENTIFIED A DESIGN DEFICIENCY AT 4D INTEGRATED TREATMENT CONSOLE: INCORRECT DISPLAY OF A FIELD CIAO (COMPLETE IRRADIATED AREA OUTLINE) IN 4DITC. IT ONLY HAPPENS IN CASE OF AD-HOC IMAGING. THE 4DITC'S CALCULATION FOR AD-HOC IMAGE TEMPLATE - THIS ISSUE IS RELATED TO INCORRECT CIAO DISPLAYED FOR IMAGING FIELD AND IN TURN THIS MAKES MLC LEAF POSITIONS GO AS PER THE CIAO AND IN TURN THIS MAKES THE IMAGE TAKEN AS PER CIAO. IF ELECTRON FIELDS ARE ABUTTED TO CIAO, THE RESULTING ELECTRON FIELD WOULD BE TOO SMALL, WHICH WILL LEAD TO UNDER DOSE OF TARGET VOLUME. STANDARD RADIATION THERAPY PRACTICE INCLUDES ROUTINE PT SETUP AND PLANNED ISOCENTER/TREATMENT VOLUME VERIFICATION PRIOR TO TREATMENT DELIVERY REGARDLESS OF DOCUMENTED PT PLAN SETUP INSTRUCTIONS. THIS VERIFICATION CAN BE DONE THROUGH THE USE OF MV, KV IMAGING OR USE OF FILM PRIOR TO PT TREATMENT. IN ADDITION, CLINICAL VISUALIZATION OF THE ELECTRON BLOCK MATCH - USER WOULD HAVE TO GO INSIDE ROOM AND PERFORM MANUAL MATCH. THIS USE CASE IS LIMITED TO ONLY AD-HOC IMAGING AND ELECTRON ABUTTING TREATMENT. NO PT INJURY REPORTED. A CUSTOMER TECHNICAL BULLETIN WILL BE SENT TO AFFECTED CUSTOMERS TO PROVIDE CLARIFICATION AND FURTHER INSTRUCTIONS REGARDING THIS ISSUE. NO FOLLOW-UP REPORTS ARE ANTICIPATED.
MIDLINE BAR ON FIELD APPEARS INCORRECT WHEN IMAGED. THE FOLLOWING WAS REPORTED: MLC GRAPHICAL DISPLAY IS CORRECT WHEN FIELD IS MODED UP FOR TREATMENT. UPON ADDING A "BEFORE PLANNED" IMAGE TEMPLATE, THE MLC FIELD GRAPHIC DISPLAY IS NO LONGER ACCURATE. IT APPEARS A PORTION OF THE MLB IS MISSING AND C-SPINE IS NOT BLOCKED AS PLANNED. THE CUSTOMER VERIFIED THAT THE ANOMALOUS DISPLAY OCCURS ONLY DURING IMAGING. TREATMENT IS NOT AFFECTED. THERE WAS NO REPORT OF INJURY TO THE PT AND NO PT DATA WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 4D INTEGRATED TREATMENT CONSOLE | LINEAR ACCELERATOR SOFTWARE | IYE | VARIAN MEDICAL SYSTEMS | H51 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |