FDA Adverse Event Malfunction Summary report: N

TUBING SETS

MDR report key: 18759082 · Received February 22, 2024

Report

Report Number
8010762-2024-00098
Event Type
Malfunction
Date Received
February 22, 2024
Date of Event
February 7, 2024
Report Date
March 8, 2024
Manufacturer
MAQUET CARDIOPULMONARY GMBH
Product Code
DWE
PMA / PMN Number
K080592
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

IT WAS REPORTED THAT HOLE WAS FOUND ON TYVEK WHILE PREPARING THE EQUIPMENT. THEREFORE, THE PRODUCT WAS NOT USED FOR TREATMENT. THE PACK IS QUARANTINED. NO HARM TO ANY PERSON HAS BEEN REPORTED. QH 2023.02 WAS INITIATED FOR THE PRODUCT ¿BE-MECC 101403 / 701075208, V04¿. FURTHER, A CAPA HAS BEEN INITIATED FOR THE REPORTED FAILURE ¿HOLE AT TYVEK / TYVEK PERFORATED¿ AND FOR THE REPORTED PRODUCT ¿BE-MECC 101403 / 701075208¿. THE ROOT CAUSE ANALYSIS AND FURTHER ACTIONS TO DETERMINE CORRECTIVE MEASURES FOR THE FAILURE WILL BE PERFORMED WITHIN THE CAPA. ALL FURTHER STEPS WILL BE PERFORMED IN ACCORDANCE TO CAPA. THE PRODUCTION HISTORY RECORDS (DHRS) OF THE AFFECTED BE-MECC 101403 WITH LOT# 3000315701 WAS ALSO REVIEWED ON 2024-02-12. ACCORDING TO THE DHR RESULTS, THE PRODUCT BE-MECC 101403 PASSED THE DEFINED MANUFACTURING AND FINAL RELEASE SPECIFICATIONS. BASED ON THE RESULTS, AND PROVIDED PHOTOGRAPHICAL EVIDENCE, THE REPORTED FAILURE "HOLE (PERFORATION) AT TYVEK" COULD BE CONFIRMED. THE OCCURRENCE RATE RELATED TO THE REPORTED ISSUE IS CURRENTLY BEING MONITORED AS PART OF MAQUET CARDIOPULMONARY¿S TRENDING PROGRAM AND ADDITIONAL INVESTIGATIONS OR CORRECTIONS WILL BE IMPLEMENTED IN CASE OF ADVERSE TRENDING.

Additional Manufacturer Narrative · 0

A FOLLOW-UP MEDWATCH WILL BE SUBMITTED WHEN FURTHER INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

COMPLAINT # (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT HOLE WAS FOUND ON TYVEK WHILE PREPARING THE EQUIPMENT. THEREFORE, THE PRODUCT WAS NOT USED FOR TREATMENT. NO HARM TO ANY PERSON WAS REPORTED. COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1642948 TUBING SETS TUBING, PUMP, CARDIOPULMONARY BYPASS DWE MAQUET CARDIOPULMONARY GMBH BE-MECC 101403 3000315701

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other