FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP

MDR report key: 1875366 · Received October 19, 2010

Report

Report Number
6000001-2010-04304
Event Type
Malfunction
Date Received
October 19, 2010
Date of Event
March 27, 2010
Report Date
September 28, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED AND DUPLICATED. THE ASSIGNABLE CAUSE WAS A FAULTY PHM (PUMPHEAD MODULE). THE PHM FOR CHANNEL A HAS BEEN REPLACED. ADDITIONAL: A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS NOT PREVIOUSLY SERVICED PRIOR TO THIS EVENT. A DEVICE HISTORY REVIEW WAS PERFORMED AND THERE WERE NO EXCEPTIONS TO THE MANUFACTURING OF THIS DEVICE. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED CONDITION. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

DURING A REVIEW OF THE EVENT HISTORY, IT WAS DISCOVERED THAT FAILURE CODE 808:02 OCCURRED ONCE DURING INFUSION AND CAUSED AN INTERRUPTION DURING DELIVERY. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION FOR THIS DEVICE IS 5.04.00, CATEGORIZED AS UNREMEDIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1