VIRTUOSO VR
Report
- Report Number
- 6000094-2010-02065
- Event Type
- Injury
- Date Received
- October 19, 2010
- Date of Event
- June 24, 2010
- Manufacturer
- MEDTRONIC S.A.
- Product Code
- LWS
- PMA / PMN Number
- ASKU
- Removal / Correction Number
- Z-0131-2010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY (B)(4) A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS FOUND THE CAUSE OF THE HIGH CURRENT DRAIN TO BE CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.
IT WAS REPORTED THAT THE DEVICE WAS REPLACED DUE TO BATTERY DEPLETION. DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING ANALYSIS. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VIRTUOSO VR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC S.A. | D164VWC | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 | Other |