FDA Adverse Event Malfunction Summary report: N

SYRINGE 20ML LL S/C 48

MDR report key: 18752881 · Received February 21, 2024

Report

Report Number
1911916-2024-00122
Event Type
Malfunction
Date Received
February 21, 2024
Date of Event
February 14, 2024
Report Date
March 1, 2024
Manufacturer
BECTON DICKINSON
Product Code
FMF
UDI-DI
30382903028307
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4) FOLLOW UP FOR DEVICE EVALUATION. IT WAS REPORTED THERE WAS FOREIGN MATTER BETWEEN THE SYRINGE AND THE PLUNGER. TO AID IN THE INVESTIGATION, ONE SAMPLE IN AN OPENED PACKAGING BLISTER AND ONE PHOTO WAS RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. A VISUAL INSPECTION WAS PERFORMED AND THERE IS A STRING-LIKE PLASTIC FLASH OF BURNT PLASTIC ADHERED TO THE PLUNGER ROD. IT IS NOT LOOSE AND LOCATED AT THE BOTTOM OF THE PLUNGER ROD, RIGHT ABOVE THE RUBBER STOPPER. THE PHOTO PROVIDED SHOWS THE SAMPLE RECEIVED. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. THIS DEFECT COULD OCCUR DURING THE MOLDING PROCESS. DEGRADED RESIN INHERENTLY BUILDS UP IN THE SYSTEM OF THE TOOLING MOLD AND PRESS. THE DEGRADED RESIN CAN BREAK LOOSE AND BE MOLDED INTO COMPONENTS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 302830, LOT 2210055. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE RETURNED SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED.

Additional Manufacturer Narrative · 0

(B)(4): INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED THERE IS A FOREIGN MATERIAL BETWEEN THE SYRINGE AND THE PLUNGER.

Description of Event or Problem · 0

THERE IS A FOREIGN MATERIAL BETWEEN THE SYRINGE AND THE PLUNGER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1484645 SYRINGE 20ML LL S/C 48 SYRINGE, PISTON FMF BECTON DICKINSON 2210055 30382903028307

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown