CONCERTO CRT-D DR
MDR report key: 1875275
·
Received October 19, 2010
Report
- Report Number
- 6000144-2010-05467
- Event Type
- Injury
- Date Received
- October 19, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P010031/S031
- Removal / Correction Number
- Z-0126-2010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS FOUND THE CAUSE OF THE HIGH CURRENT DRAIN TO BE CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.
Description of Event or Problem · 1
THE DEVICE WAS REPLACED BECAUSE IT WAS AT ERI (ELECTIVE REPLACEMENT INDICATOR) SOONER THAN ANTICIPATED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.