FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CXE VOLUMETRICINFUSION PUMP

MDR report key: 1875260 · Received October 19, 2010

Report

Report Number
6000001-2010-04300
Event Type
Malfunction
Date Received
October 19, 2010
Date of Event
September 10, 2010
Report Date
September 29, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA-(B)(4). THE DEVICE HAS BEEN PREVIOUSLY SENT INTO SERVICE FOR THE REPORTED CONDITION OF FC 810:11. (B)(4)

Additional Manufacturer Narrative · 1

THE REPORTED CONDITION OF FAILURE CODE 810:11 WAS CONFIRMED THROUGH THE EVENT HISTORY, BUT NOT DUPLICATED. THE REPORTED CONDITION IS DUE TO THE AIL PCB (AIR-IN-LINE PRINTED CIRCUIT BOARD) BEING OUT OF CALIBRATION. THE AIL PCB WAS RECALIBRATED TO CORRECT THE REPORTED CONDITION. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4).

Description of Event or Problem · 1

DURING BAXTER'S REVIEW OF THE EVENT HISTORY, IT WAS DISCOVERED THAT FAILURE CODE 810:11 OCCURRED ON (B)(6) 2010 DURING DELIVERY CAUSING AN INTERRUPTION OF DELIVERY. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THIS DEVICE. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.90 CATEGORIZED AS REMEDIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CXE VOLUMETRICINFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1