Narratives
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ACTUAL LONGEVITY IS < 80% OF 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.
IT WAS REPORTED THAT THE DEVICE WAS REPLACED AS IT HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). THE DEVICE TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.