FDA Adverse Event Malfunction Summary report: N

6461500 INFUSOR PUMP (1 LABEL)

MDR report key: 1875248 · Received October 19, 2010

Report

Report Number
6000001-2010-04293
Event Type
Malfunction
Date Received
October 19, 2010
Date of Event
September 24, 2010
Report Date
September 27, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K883577
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) THE DEVICE EVALUATION IS IN PROGRESS. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED AND DUPLICATED. THE ASSIGNABLE CAUSE WAS A FAULTY MPU (MICROPROCESSING UNIT) BOARD. THE MPU BOARD HAS BEEN REPLACED. ADDITIONAL: A SERVICE HISTORY REVIEW HAS BEEN PERFORMED. THE DEVICE HAS NOT BEEN SERVICED BY BAXTER PRIOR TO THIS EVENT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED CONDITION. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT AN INFUSOR THAT HAD INFUSION ALARMS OCCUR DURING AN ACCURACY TEST. THE EVENT OCCURRED DURING BIOMEDICAL TESTING. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. DURING INITIAL PRODUCT EVALUATION AT BAXTER, IT WAS DISCOVERED THAT AFTER SHORTLY STARTING TO INFUSE, THE DEVICE WENT INTO CONSTANT ALARM WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6461500 INFUSOR PUMP (1 LABEL) PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1