FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1875195 · Received October 19, 2010

Report

Report Number
1423500-2010-04633
Event Type
Malfunction
Date Received
October 19, 2010
Date of Event
September 25, 2010
Report Date
September 25, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS DISCARDED. SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). AN ENGINEERING QUALITY REVIEW WAS COMPLETED FOR THIS REPORT OF A SYSTEM ERROR (SE) 2240 ALARM. THE REPORT WAS NOT CONFIRMED DUE TO LACK OF SAMPLE. THE LOT NUMBER WAS UNKNOWN; THEREFORE, A BATCH REVIEW WAS NOT PERFORMED. BASED ON THE INFORMATION OBTAINED FROM BAXTER'S INVESTIGATION, THE ROOT CAUSE OF THE SE 2240 WAS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE BECAUSE THE PATIENT EXTENSION LINE WAS CONNECTED AFTER PRIMING WAS COMPLETE. THE CAREGIVER (CG) STATED THAT SHE FORGOT TO HOOK THE PATIENT EXTENSION LINE BEFORE PRIMING. THE LABELING REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

A CAREGIVER (CG) CONTACTED GLOBAL TECHNICAL SERVICES REGARDING A SYSTEM ERROR (SE) 2240 ALARM THAT OCCURRED ON THE HOMECHOICE (HC) UNIT DURING INITIAL DRAIN. THE CG STATED THAT SHE FORGOT TO HOOK THE PATIENT EXTENSION LINE BEFORE PRIMING. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE CG TO CYCLE POWER TO THE HC TO CLEAR THE ALARM. THE CG CONFIRMED TO SETUP WITH NEW SUPPLIES. THE TSR REVIEWED PROPER PROCEDURES PER THE USER MANUAL WITH THE CG. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 76 YR