FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 1875173 · Received October 19, 2010

Report

Report Number
2124215-2010-18421
Event Type
Malfunction
Date Received
October 19, 2010
Date of Event
September 9, 2010
Report Date
October 19, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Removal / Correction Number
Z-0863-05
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WERE ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED INDICATING THE DEVICE WAS EXPLANTED. THE DEVICE WAS NEAR ERI AND WAS ELECTIVELY REPLACED. THE DEVICE WAS RETURNED FOR ANALYSIS. THIS EVENT WILL BE UPDATED WHEN ANALYSIS IS COMPLETE.

Additional Manufacturer Narrative · 1

A DEVICE REPLACEMENT AND LV LEAD REVISION WERE PLANNED FOR A FUTURE DATE. THE AVAILABLE INFORMATION SUGGESTS THIS DEVICE REMAINS IN SERVICE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS EVENT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). LEFT VENTRICULAR (LV) PACING IMPEDANCE MEASUREMENTS GREATER THAN 2000 OHMS WERE OBSERVED AS WELL AS LOSS OF CAPTURE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H217

Patients

Seq Age Sex Outcome Treatment
1 76 YR H217| 0180| 4542| 4469