FDA Adverse Event Malfunction Summary report: N

ENTRUST AT

MDR report key: 1875141 · Received October 19, 2010

Report

Report Number
6000144-2010-05446
Event Type
Malfunction
Date Received
October 19, 2010
Date of Event
November 23, 2009
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S46
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. CRITICAL RAM PARITY ERROR NOTED ON (B)(6) 2009. POR (POWER ON RESET) SEVERITY: LOW. THE DEVICE SHOULD BE ABLE TO FULLY RECOVER AFTER THE RESET.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CORRECTION: CHANGE MADE TO DEVICE CONCLUSION. DEVICE CODE CHANGED. EVALUATION SUMMARY: (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. CRITICAL RAM PARITY ERROR NOTED ON (B)(6) 2009. POR (POWER ON RESET) SEVERITY: LOW. THE DEVICE SHOULD BE ABLE TO FULLY RECOVER AFTER THE RESET.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE EXPERIENCED AN ELECTRICAL RESET. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTRUST AT IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D154ATG ASKU

Patients

Seq Age Sex Outcome Treatment
1 58 YR Other 6943 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD| 6943 IMPLANTABLE TACHY LEAD