CONCERTO CRT-D DR

MDR report key: 1875135 · Received October 19, 2010

Report

Report Number
6000144-2010-05448
Event Type
Injury
Date Received
October 19, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S031
Removal / Correction Number
Z-0126-2010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS REPLACED DUE TO REACHING ELECTIVE REPLACEMENT INDICATOR (ERI). THE DEVICE TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO CRT-D DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. C154DWK ASKU

Patients

Seq Age Sex Outcome Treatment
1 78 YR Required Intervention 4076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD