INFUSOR LV5, 12 PK
Report
- Report Number
- 6000001-2010-04271
- Event Type
- Malfunction
- Date Received
- October 19, 2010
- Date of Event
- September 24, 2010
- Report Date
- September 24, 2010
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: ONE SAMPLE WAS RECEIVED BY BAXTER FOR EVALUATION. THE REPORTED CONDITION OF "BACKFLOW" WAS CONFIRMED DUE TO A GLASS FRAGMENT (APPROXIMATELY 0.9 MM IN SIZE) FOUND TRAPPED BETWEEN THE CHECK-BAND AND STRESS-MEMBER. THIS IS A SINGLE USE DEVICE AND WILL NOT BE REPAIRED. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT.
(B)(4). ADDITIONAL NARRATIVE: PER THE CUSTOMER, THE DEVICE IS AVAILABLE FOR EVALUATION; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE AND/OR ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.
IT WAS REPORTED TO BAXTER (B)(4) THAT ONE (1) INFUSOR LV5 DEVICE WAS OBSERVED BACKFLOWING FROM THE FILL PORT DURING FILLING. THE DEVICE WAS BEING FILLED WITH 5-FLUOROURACIL AND NORMAL SALINE. THERE IS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.