FDA Adverse Event Injury Summary report: N

ADAPTA SR

MDR report key: 1875007 · Received October 19, 2010

Report

Report Number
6000144-2010-05438
Event Type
Injury
Date Received
October 19, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NVZ
PMA / PMN Number
P980035/S043
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS ANALYZED AND DID NOT MEET ITS EXPECTED LONGEVITY. THE CAUSE IS UNKNOWN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS POSSIBLE PREMATURE BATTERY DEPLETION, BUT THAT DEVICE EXPERIENCED HIGH OUTPUT DUE TO INCREASED RV THRESHOLDS ON A COMPETITOR LEAD. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADAPTA SR IMPLANTABLE PULSE GENERATOR NVZ MEDTRONIC MED REL, INC. ADSR01 ASKU

Patients

Seq Age Sex Outcome Treatment
1 86 YR Other| R (B)(4) COMPETITOR IMPLANTABLE PACING LEAD