FDA Adverse Event Malfunction Summary report: N

BD INSYTE AUTOG BC BLU 22GA X 1.0IN

MDR report key: 18749474 · Received February 21, 2024

Report

Report Number
1710034-2024-00119
Event Type
Malfunction
Date Received
February 21, 2024
Date of Event
November 16, 2023
Report Date
March 20, 2024
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903825233
PMA / PMN Number
K201075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. A.2. DATE OF BIRTH: UNKNOWN. THE PATIENT¿S AGE WAS USED TO DETERMINE A PLACEHOLDER DATE FOR THIS FIELD. E4. THE INITIAL REPORTER ALSO NOTIFIED THE FDA DECEMBER OF 2023. MEDWATCH REPORT IS (B)(4).

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 382523 AND LOT NUMBER 3192172. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INSYTE AUTOG BC BLU 22GA X 1.0IN CATHETER SHEARED. THE FOLLOWING INFORMATION WAS PROVIDED VIA MEDWATCH: REGISTERED NURSE (RN) STARTED A 22-GAUGE X 1 INCH IN THE LEFT ACROMIOCLAVICULAR JOINT. BLOOD RETURN NOTED. I TRIED TO ADVANCE CATHETER AND MET RESISTANCE. I NOTED THAT THERE WAS A "BUBBLE" WHERE THE CATHETER WAS UNDER THE SKIN. I PULLED THE NEEDLE WITH CATHETER OUT AND APPLIED PRESSURE. THE CATHETER WAS INTACT BUT SHEARED FROM NEEDLE.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1602930 BD INSYTE AUTOG BC BLU 22GA X 1.0IN INTRAVASCULAR CATHETER FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 3192172 00382903825233

Patients

Seq Age Sex Outcome Treatment
1 12 YR Female