FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1874832 · Received October 19, 2010

Report

Report Number
2649622-2010-11579
Event Type
Injury
Date Received
October 19, 2010
Date of Event
August 12, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD DISLODGED. THE LEAD WAS EXTRACTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 50 YR Required Intervention 6944 IMPLANTABLE TACHY LEAD| 4196 IMPLANTABLE PACING LEAD| D224TRK IMPLANTABLE PACEMAKER/CARDIO/DEFIB