FDA Adverse Event Malfunction Summary report: N

REGARD

MDR report key: 18748199 · Received February 21, 2024

Report

Report Number
3014527682-2024-00006
Event Type
Malfunction
Date Received
February 21, 2024
Report Date
February 21, 2024
Manufacturer
ROI CPS, LLC
Product Code
LRO
UDI-DI
10194717116202
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

A SITE VISIT TO THE FACILITY WHO REPORTED THIS EVENT WAS CONDUCTED AFTER ROI CPS, LLC IDENTIFIED THAT THE KIT LOT REPORTED IN THIS COMPLAINT HAD BEEN 100% INSPECTED FOR DAMAGES PRIOR TO SHIPMENT. THE INSPECTION CONDUCTED BY OUR FIRM CONFIRMED THERE WERE NO DAMAGES TO THE VENTED BAGS PRIOR TO THE KITS LEAVING OUR FIRM AND ARRIVING AT THE END USER FACILITY. DURING THE SITE VISIT, ROI CPS, LLC KITS WERE OBSERVED FROM RECEIVING, THROUGH HANDLING, STORING, AND USE OF THE KITS, WITH END USER FACILITY STAFF PRESENT. THE CONCLUSIONS DRAWN FROM THE VISIT AND THE RECOMMENDATIONS TO THE FACILITY ARE ATTACHED.

Additional Manufacturer Narrative · 0

A SITE VISIT TO THE FACILITY WHO REPORTED THIS EVENT WAS CONDUCTED AFTER ROI CPS, LLC IDENTIFIED THAT THE KIT LOT REPORTED IN THIS COMPLAINT HAD BEEN 100% INSPECTED FOR DAMAGES PRIOR TO SHIPMENT. THE INSPECTION CONDUCTED BY OUR FIRM CONFIRMED THERE WERE NO DAMAGES TO THE VENTED BAGS PRIOR TO THE KITS LEAVING OUR FIRM AND ARRIVING AT THE END USER FACILITY. DURING THE SITE VISIT, ROI CPS, LLC KITS WERE OBSERVED FROM RECEIVING, THROUGH HANDLING, STORING, AND USE OF THE KITS, WITH END USER FACILITY STAFF PRESENT. THE CONCLUSIONS DRAWN FROM THE VISIT AND THE RECOMMENDATIONS TO THE FACILITY ARE ATTACHED. ON 4/19/2024 ROI CPS, LLC RECEIVED AN EMAIL COMMUNICATION FROM THE USER FACILITY WHERE THIS KIT IS STORED THAT OUTLINED THEIR FINDINGS AT THE FACILITY RELATED TO THE ROOT CAUSE OF THE DAMAGE TO KITS STORED AT THE USER FACILITY. THE COMMUNICATION FROM THE FACILITY IS AS FOLLOWS: WE MAY HAVE DISCOVERED A SOURCE OF THE SNAGS ON THE MINOR PACKS. ALONG THE OUTER EDGE OF THE METRO SHELVES HOUSING THE MINOR PACKS IS A BLUE STRIP THAT IS FOR LABELS/ TAGS TO BE ATTACHED TO THE SHELF. WE DON'T USE THEM HOWEVER THREE OF THE APPROX. 10 CARTS IN THE CASE CART AREA HAVE THESE STRIPS. TWO OF THE CARTS CONTAIN NON PACK ITEMS THAT WOULD NOT BE IN CONTACT WITH THESE STRIPS. THE ONLY PACKS TO BE STOCKED ON THESE PARTICULAR CARTS ARE MINOR PACKS THAT ARE HOUSED ON ONE CART. BY HAPPEN STANCE ONE OF THE PACKS GOT NOTICEABLY SNAGGED BY THIS SHELF STRIP. IT MADE A LOUD NOISE THAT CAUGHT THE ATTENTION OF THE CASE CART BUILDER. IN CLOSER EXAMINATION THE BOTTOM TWO SHELF STRIPS WERE RATHER JAGGED WHEN YOU RAN YOUR FINGER OVER IT. BECAUSE IT WAS THE BOTTOM TWO SHELFS IT WAS NOT VISUALLY NOTICEABLE. HOWEVER, IT WAS PRONOUNCED ENOUGH THAT I THINK IT MAY BE THE SOURCE OF THE PROBLEM. 99% OF THE ISSUES WE ARE HAVING ARE MINOR PACKS AND THIS DISCOVERY IS THE ONLY DIFFERENTIATOR THAT I CAN SEE BETWEEN MINOR AND OTHER PACKS. I THINK HAVING THE BASIN IN THE PACK HAS MADE IT MORE VULNERABLE TO THE ROUGH EDGES LIKE THE ONES ON THESE TWO SHELVES. STRIPS HAVE BEEN REMOVED.

Description of Event or Problem · 0

A HOLE WAS DISCOVERED IN THE VENTED BAG.

Description of Event or Problem · 0

A HOLE WAS DISCOVERED IN THE VENTED BAG.

Description of Event or Problem · 0

A HOLE WAS DISCOVERED IN THE VENTED BAG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1444158 REGARD MINOR PACK LRO ROI CPS, LLC GS00085AD 98644 10194717116202

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown