FDA Adverse Event
Injury
Summary report: N
MEGA SOFT REUSABLE PATIENT RETURN ELECTRODE
MDR report key: 18747981
·
Received February 21, 2024
Report
- Report Number
- 18747981
- Event Type
- Injury
- Date Received
- February 21, 2024
- Date of Event
- January 2, 2024
- Report Date
- January 5, 2024
- Manufacturer
- MEGADYNE MEDICAL PRODUCTS
- Product Code
- GEI
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- MI
- Reporter Occupation
- RISK MANAGER
- Health Professional
- N
Narratives
Description of Event or Problem · 0
YOUNG CHILD HAVING ADENOIDECTOMY SURGERY. OPERATIVE TABLE CONSISTED OF GELLI-ROLL HEATING PAD WITH MEGADYNE PATIENT RETURN ELECTRODE PAD ATOP AND THEN A SAHARA SUPER-ABSORBENT OPERATING ROOM (OR) TABLE SHEET, IMPERVIOUS, ON TOP OF THAT WITH WHITE, CLOTH SIDE TO PATIENT. PATIENT THEN HAD A DRAW SHEET UNDER HIM, ABOVE THE OR TABLE SHEET. FOLLOWING THE 11 MINUTE SURGERY, HE WAS BROUGHT TO THE PEDIATRIC OUTPATIENT SURGERY AREA TO RECOVER, HOWEVER, THE REGISTERED NURSE (RN) NOTICED A BURN TO THE PATIENT'S RIGHT BUTTOCK. PATIENT HAS BEEN REFERRED TO A BURN CENTER FOR EVALUATION AND TREATMENT. HE MAY NEED PLASTIC SURGERY IN HIS FUTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 47899 | MEGA SOFT REUSABLE PATIENT RETURN ELECTRODE | ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES | GEI | MEGADYNE MEDICAL PRODUCTS | 0845 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 YR | Male | Other| R| H |