FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1874759 · Received October 19, 2010

Report

Report Number
2649622-2010-11545
Event Type
Injury
Date Received
October 19, 2010
Date of Event
August 27, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND THE PROXIMAL CONDUCTOR FRACTURED. IT WAS NOTED THAT THE OUTER TUBING WAS KINKED/BUCKLED, THE OUTER TUBING OVERLAY HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER TUBING OVERLAY WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR WAS DISTORTED. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THE INNER TUBING WAS KINKED/BUCKLED, THE HELIX/LOBE WAS DISTORTED/BENT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM (SLEEVE HEAD), AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. DAMAGED AT IMPLANT. IT WAS NOTED THE LEAD WAS RECEIVED WITH THE DISTAL COIL DISTORTED WITHIN THE CONNECTOR. THE HELIX WILL HELIX WILL NOT EXTEND OR RETRACT DUE TO THE DISTAL CONDUCTOR DISTORTION.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD FAILED. DURING THE EXTRACTION PROCEDURE, A NEW RV LEAD WAS ATTEMPTED TO BE IMPLANTED, BUT WAS UNSUCCESSFUL DUE TO THE HELIX NOT EXTENDING OR RETRACTING. ANOTHER LEAD WAS SUCCESSFULLY IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RV LEAD FAILED. DURING THE EXTRACTION PROCEDURE, A NEW RV LEAD WAS ATTEMPTED TO BE IMPLANTED, BUT WAS UNSUCCESSFUL DUE TO THE HELIX NOT EXTENDING OR RETRACTING. ANOTHER LEAD WAS SUCCESSFULLY IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 84 YR Hospitalization| R 4194 IMPLANTABLE PACING LEAD| C154DWK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5568 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| C154DWK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5568 IMPLANTABLE PACING LEAD