VITALITY 2
Report
- Report Number
- 2124215-2010-17761
- Event Type
- Injury
- Date Received
- October 19, 2010
- Date of Event
- September 8, 2010
- Report Date
- October 4, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, DEVICE MEMORY WAS REVIEWED. WE CONFIRMED THAT THE DEVICE DECLARED ERI AFTER EXPERIENCING ONE CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT OF 18.9 SECONDS. TO DETERMINE IF THE DEVICE'S RATE OF BATTERY DEPLETION WAS NORMAL, OUR TECHNICIANS COMPARED THE OBSERVED RATE OF BATTERY USAGE TO THE EXPECTED RATE OF BATTERY USAGE. THE RESULTS SHOWED THAT THE MONITORING VOLTAGE WAS NORMAL BASED ON THERAPY USE TO DATE AND PROGRAMMED SETTINGS. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY CHARGE TIMES IN EXCESS OF THE 18.9 SECOND EXTENDED ERI. LABORATORY TECHNICIANS AND ENGINEERS CONCLUDED THAT THIS DEVICE DID NOT EXPERIENCE PREMATURE BATTERY DEPLETION. RATHER, AN ERI CHARGE TIME REPLACEMENT INDICATOR WAS DECLARED DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.
UPON ANALYSIS THIS EVENT WILL BE UPDATED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED AFTER THE DEVICE TRIGGERED ELECTIVE REPLACEMENT INDICATOR (ERI). PREMATURE BATTERY DEPLETION WAS SUSPECTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T177 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |