FDA Adverse Event Injury Summary report: N

VIRTUOSO II DR

MDR report key: 1874709 · Received October 19, 2010

Report

Report Number
6000144-2010-05412
Event Type
Injury
Date Received
October 19, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S157
Removal / Correction Number
ASKU
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS POCKET EROSION AND INFECTION. THE DEVICE WAS REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO II DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D274DRG ASKU

Patients

Seq Age Sex Outcome Treatment
1 92 YR Required Intervention 5076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD