FDA Adverse Event Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1874688 · Received October 15, 2010

Report

Report Number
2531779-2010-01682
Date Received
October 15, 2010
Date of Event
September 15, 2010
Report Date
September 16, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED THAT ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS AFTER THE KEYPAD WAS REMOVED. CONTAMINATION WAS OBSERVED UNDER THE BUTTONS DURING INVESTIGATION. THE KEYPAD WAS PEELING AT THE OK BUTTON. THE BUTTONS WERE RESPONSIVE AND SPRINGING BACK NORMALLY. DATE OF BIRTH IS UNKNOWN.

Description of Event or Problem · 1

THE DISTRIBUTOR CONTACTED ANIMAS ALLEGING THAT THE PUMP WAS INTERMITTENTLY NOT RESPONDING TO BUTTON PRESSES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1