FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY

MDR report key: 1874680 · Received October 19, 2010

Report

Report Number
2649622-2010-11506
Event Type
Injury
Date Received
October 19, 2010
Date of Event
August 16, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND. ALL CONDUCTORS DISTORTED AND BLOOD/BODY FLUID (NOT OBSTRUCTED). OUTER INSULATION MELTED. FULL LEAD RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD HAD POSITIONING/FIXATION DIFFICULTY. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY IMPLANTABLE PACING LEAD OJX MEDTRONIC PUERTO RICO, INC. 4196 ASKU

Patients

Seq Age Sex Outcome Treatment
1 82 YR Required Intervention 5076 (X2) IMPLANTABLE PACING LEAD| 8042 IMPLANTABLE PULSE GENERATOR