FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 1874662
·
Received October 19, 2010
Report
- Report Number
- 2649622-2010-11487
- Event Type
- Malfunction
- Date Received
- October 19, 2010
- Date of Event
- August 24, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S017
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD WAS FOUND TO BE PROGRAMMED UNIPOLAR, AND NO LEAD WARNING WAS NOTED. ADDITIONALLY THE LEAD WAS PROGRAMMED TO BIPOLAR AND THEN THE LEAD HAD NO CAPTURE, NO SENSING, AND HAD LOW LEAD IMPEDANCE. THE UNIPOLAR IMPEDANCE WAS HIGHER THAN THE BIPOLAR IMPEDANCE. THE LEAD IS STILL IN USE. NO PATIENT COMPLICATIONS WERE REPORTED DUE TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |