CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2010-11439
- Event Type
- Malfunction
- Date Received
- October 19, 2010
- Date of Event
- August 20, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS RESULTS REVEALED THAT NO ANOMALIES WERE FOUND. BLOOD WAS PRESENT IN/ON THE HELIX MECHANISM.
IT WAS REPORTED THAT THE LEAD WAS IMPLANTED AND AFTER SEVERAL REPOSITIONING ATTEMPTS LOW P-WAVE AND IMPEDANCE WERE MEASURED. IT WAS NOTICED THAT THE STYLET WAS NOT EASY TO INSERT/RETRACT. THE HELIX MECHANISM OF THE LEAD DIDN'T WORK PROPERLY. THE LEAD WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |