Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND. A WHITE SUBSTANCE WAS NOTED ON THE LEAD. THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD HAD HIGH IMPEDANCE, UNSTABLE THRESHOLDS AND SENSING DIFFICULTY. THE LEAD WAS PARTIALLY REMOVED AND FOUND TO BE FRACTURED. A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 98 YR Required Intervention 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR