FDA Adverse Event Malfunction Summary report: N

FLO-GARD 6201 VOLUMETRIC INFUSION PUMP

MDR report key: 1874477 · Received October 19, 2010

Report

Report Number
6000001-2010-04266
Event Type
Malfunction
Date Received
October 19, 2010
Date of Event
September 1, 2010
Report Date
September 24, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH FAILURE CODE 2 WAS CONFIRMED AND REPRODUCED DURING PRODUCT EVALUATION. THIS CONDITION WAS CAUSED BY THE PUMP'S DOOR ASSEMBLY BEING CRACKED AT THE LOWER HINGE STOP. THE DOOR ASSEMBLY WAS REPLACED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW REVEALED THE DEVICE HAS BEEN PREVIOUSLY SENT INTO SERVICE FOR THE REPORTED CONDITION OF F-2. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

THE FACILITY REPORTED A FLO-GARD INFUSION PUMP WITH FAILURE CODE 2, WHICH IS A DOWNSTREAM OCCLUSION FAILURE CODE. ACCORDING TO THE FACILITY, IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. THE FACILITY REPRESENTATIVE DID NOT HAVE INFORMATION REGARDING WHETHER THERE HAVE BEEN ANY REPORTS OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. THIS FACILITY WAS NOT WILLING TO BE CONTACTED FOR ANY FURTHER INFORMATION. DURING PRODUCT EVALUATION, FAILURE CODE 2 WAS CONFIRMED TO BE CAUSED BY THE DOOR ASSEMBLY BEING CRACKED AT THE LOWER HINGE STOP, INDICATING FAILURE CODE 2 WAS A FALSE OCCLUSION ALARM. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLO-GARD 6201 VOLUMETRIC INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1