FDA Adverse Event Injury Summary report: N

DEXTRUS 4137

MDR report key: 1874468 · Received October 14, 2010

Report

Report Number
1028232-2010-02148
Event Type
Injury
Date Received
October 14, 2010
Date of Event
July 28, 2010
Report Date
September 13, 2010
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVZ
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT FOUR MONTHS POST IMPLANT, THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOSS OF CAPTURE. THE X-RAY DID NOT SHOW A DISLODGEMENT, HOWEVER, THERE WAS CONCERN OVER A MICRO-DISLODGEMENT. THE RV LEAD WAS SUCCESSFULLY REPOSITIONED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. AS THE PRODUCT REMAINS IN SERVICE, IT WILL NOT BE RETURNED FOR ANALYSIS. THERE IS NO ADDITIONAL INFORMATION RELATED TO THIS EVENT AVAILABLE TO THE COMPANY AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THE EVENT WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEXTRUS 4137 PACER LEAD NVZ BIOTRONIK SE & CO. KG 358764

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization