FDA Adverse Event Injury Summary report: N

DEXTRUS 4135

MDR report key: 1874425 · Received October 14, 2010

Report

Report Number
1028232-2010-02140
Event Type
Injury
Date Received
October 14, 2010
Date of Event
August 4, 2010
Report Date
September 13, 2010
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVZ
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT FOUR DAYS POST IMPLANT, AN X-RAY REVEALED THAT THIS LEAD HAD DISLODGED. DURING THE LEAD REVISION PROCEDURE, THE FLUOROSCOPY IMAGE SHOWED THE LEAD HAD NOT BEEN ANCHORED DOWN PROPERLY AND WERE SLIDING BACK AND FORTH. THE LEAD WAS SUCCESSFULLY REPOSITIONED WITH GOOD MEASUREMENTS AND REMAINS IMPLANTED IN THE PATIENT. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS EVENT WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEXTRUS 4135 PACER LEAD NVZ BIOTRONIK SE & CO. KG 358744

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization