FDA Adverse Event Injury Summary report: N

PUMP MMT-722NAS PRDGM INS V2.2 SK EN

MDR report key: 1874205 · Received October 14, 2010

Report

Report Number
3004209178-2010-83099
Event Type
Injury
Date Received
October 14, 2010
Date of Event
October 2, 2010
Report Date
October 3, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED FOR HIGH BLOOD GLUCOSE. THE BLOOD GLUCOSE READING AT TIME OF THE CALL WAS 485MG/DL. TROUBLESHOOTING WAS PERFORMED. THE TIME AND DATE WERE CORRECT. THE DAILY TOTALS MATCHED THE BASAL AND BOLUS. RAN A FIXED PRIME TEST AND THE INSULIN EXITED. PERFORMED A HIGH PRESSURE TEST AND THE INSULIN PUMP PASSED THE TEST. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 66 YR Hospitalization