FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1874156
·
Received October 15, 2010
Report
- Report Number
- 2531779-2010-01701
- Event Type
- Malfunction
- Date Received
- October 15, 2010
- Date of Event
- September 16, 2010
- Report Date
- September 16, 2010
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED THAT THE PUMP WAS INTERMITTENTLY RESPONDING TO KEYPAD BUTTON PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS. CONTAMINATION WAS OBSERVED UNDER BUTTONS DURING INVESTIGATION.
Description of Event or Problem · 1
THE PT CONTACTED ANIMAS ALLEGING THAT THE PUMP WAS INTERMITTENTLY NOT RESPONDING TO PRESSING OF THE OK BUTTON. THE PT ALSO CLAIMED THAT THE KEYPAD WAS PEELING. SHE DENIED THAT THE DEVICE WAS EXPOSED TO MOISTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR |