FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1874156 · Received October 15, 2010

Report

Report Number
2531779-2010-01701
Event Type
Malfunction
Date Received
October 15, 2010
Date of Event
September 16, 2010
Report Date
September 16, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED THAT THE PUMP WAS INTERMITTENTLY RESPONDING TO KEYPAD BUTTON PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS. CONTAMINATION WAS OBSERVED UNDER BUTTONS DURING INVESTIGATION.

Description of Event or Problem · 1

THE PT CONTACTED ANIMAS ALLEGING THAT THE PUMP WAS INTERMITTENTLY NOT RESPONDING TO PRESSING OF THE OK BUTTON. THE PT ALSO CLAIMED THAT THE KEYPAD WAS PEELING. SHE DENIED THAT THE DEVICE WAS EXPOSED TO MOISTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 34 YR