FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1874144 · Received October 15, 2010

Report

Report Number
2531779-2010-01697
Event Type
Malfunction
Date Received
October 15, 2010
Report Date
September 16, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE BATTERY COMPARTMENT WAS CRACKED AND THE PUMP WAS INTERMITTENTLY RESPONDING TO THE AUDIO BOLUS BUTTON. THERE WAS ALSO ADHESIVE (CONTAMINATION) FOUND ON THE BUTTON CONTACTS.

Description of Event or Problem · 1

THE PUMP IS REPORTEDLY NOT RESPONDING WHEN THE AUDIO BOLUS BUTTON IS PRESSED; HOWEVER, THE OTHER BUTTONS ARE WORKING FINE. THE PT INDICATED THAT THE KEYPAD IS NOT PEELING AND THE PUMP HAS NOT GOTTEN WET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. IR 1250 NA

Patients

Seq Age Sex Outcome Treatment
1