FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1874144
·
Received October 15, 2010
Report
- Report Number
- 2531779-2010-01697
- Event Type
- Malfunction
- Date Received
- October 15, 2010
- Report Date
- September 16, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE BATTERY COMPARTMENT WAS CRACKED AND THE PUMP WAS INTERMITTENTLY RESPONDING TO THE AUDIO BOLUS BUTTON. THERE WAS ALSO ADHESIVE (CONTAMINATION) FOUND ON THE BUTTON CONTACTS.
Description of Event or Problem · 1
THE PUMP IS REPORTEDLY NOT RESPONDING WHEN THE AUDIO BOLUS BUTTON IS PRESSED; HOWEVER, THE OTHER BUTTONS ARE WORKING FINE. THE PT INDICATED THAT THE KEYPAD IS NOT PEELING AND THE PUMP HAS NOT GOTTEN WET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | IR 1250 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |