FDA Adverse Event
Malfunction
Summary report: N
ACUITY
MDR report key: 1874113
·
Received October 19, 2010
Report
- Report Number
- 2124215-2010-18989
- Event Type
- Malfunction
- Date Received
- October 19, 2010
- Date of Event
- September 2, 2010
- Report Date
- September 2, 2010
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THIS LEAD WAS ELECTRICALLY ABANDONED. THERE IS NO FURTHER INFORMATION AVAILABLE AT THIS TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE TO OUR COMPANY, THIS EVENT WOULD BE UPDATED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THE LEFT VENTRICULAR (LV) LEAD DISLODGED. THE DEVICE WAS REPROGRAMMED TO PROVIDE ONLY RIGHT VENTRICULAR THERAPY AND THE AV DELAY WAS EXTENDED TO 240MS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACUITY | IMPLANTABLE LEAD | OJX | CPI - DEL CARIBE | 4554 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | 4470| N119| 4554| 0185 |