FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722NAP PRDGM INS V2.2 PL EN

MDR report key: 1874019 · Received October 15, 2010

Report

Report Number
2032227-2010-82925
Event Type
Malfunction
Date Received
October 15, 2010
Date of Event
September 26, 2010
Report Date
September 26, 2010
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER'S MOTHER REPORTED, A BUTTON ERROR ALARM AND A FROZEN SCREEN. TROUBLESHOOTING WAS PERFORMED, BUT THE ISSUES WERE NOT RESOLVED. ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAP PRDGM INS V2.2 PL EN INFUSION PUMP MDS MEDTRONIC MINIMED MMT-722NAP

Patients

Seq Age Sex Outcome Treatment
1 14 YR