FDA Adverse Event Malfunction Summary report: N

PUMP MMT-723NAS PRDGM INSULIN SMOKE EN

MDR report key: 1873976 · Received October 15, 2010

Report

Report Number
3004209178-2010-83136
Event Type
Malfunction
Date Received
October 15, 2010
Date of Event
September 5, 2010
Report Date
October 5, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A FROZEN DISPLAY AS A RESULT OF A FAULTY COMPONENT ON THE INTERFACE BOARD. FURTHER TESTING WAS NOT PERFORMED DUE TO THE FROZEN DISPLAY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE BATTERY WAS CHANGED TWICE AND THE INSULIN PUMP HAD AN ERROR ALARM. THE CUSTOMER STATED THAT THE ACTIVATE BUTTON WAS NOT RESPONDING, AND THE INSULIN PUMP WAS ALARMING. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-723NAS PRDGM INSULIN SMOKE EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 48 YR