FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 1873900 · Received October 19, 2010

Report

Report Number
2124215-2010-18219
Event Type
Injury
Date Received
October 19, 2010
Date of Event
September 1, 2010
Report Date
September 22, 2010
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
PMA / PMN Number
P910073
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT THE POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL ANALYSIS REVEALED THAT THE COMPLETE LEAD WAS RETURNED AND THERE WERE SETSCREW MARKS NOTED ON ALL TERMINAL CONNECTORS. THE TRILUMEN INSULATION HAD AN ABRASION 31 CENTIMETERS FROM THE IS-1 PIN. ELECTRICAL CONTINUITY TESTING WAS PERFORMED WITH MANIPULATION ON ALL CONDUCTIVE PATHWAYS. THIS LEAD WAS FOUND TO BE ELECTRICALLY CONTINUOUS. THE TRILUMEN INSULATION ABRASION WAS FURTHER ANALYZED AND THE RATE/SENSE NEGATIVE COIL AND DISTAL HIGH VOLTAGE CONDUCTORS WERE EXPOSED, THE PROXIMAL CABLE APPEARED TO BE INTACT AND NO CONDUCTORS EXPOSED. ANALYSIS REVEALED INSULATION DAMAGE AND CONDUCTOR EXPOSURE, HOWEVER, ALL CONDUCTIVE PATHS SHOW CONTINUOUS INDICATING THAT NO FRACTURE HAD OCCURRED. DUE TO LOCATION AND TYPE OF DAMAGE IT WAS DETERMINED THAT IT WAS MOST LIKELY DUE TO CLAVICLE FIRST/RIB ENTRAPMENT.

Additional Manufacturer Narrative · 1

ANALYSIS IS CURRENTLY ON-GOING. UPON COMPLETION OF ANALYSIS, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE WHICH RESULT IN INAPPROPRIATE SHOCKS ALONG WITH INCREASED PACING THRESHOLDS AND CHANGES IN PACING IMPEDANCE MEASUREMENTS. IT WAS DETERMINED THAT THE LEAD WAS FRACTURED. THE PATIENT UNDERWENT A REVISION PROCEDURE AND THIS PRODUCT WAS EXPLANTED AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0186

Patients

Seq Age Sex Outcome Treatment
1 63 YR Required Intervention H170| 4543| 4076| 0186