FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 10ML LL BNS

MDR report key: 18738843 · Received February 20, 2024

Report

Report Number
1213809-2024-00094
Event Type
Malfunction
Date Received
February 20, 2024
Date of Event
July 25, 2023
Report Date
January 30, 2024
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
50382903010290
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

IT WAS REPORTED THERE WERE BLACK MARKINGS AROUND THE SYRINGE. TO AID IN THE INVESTIGATION, FIVE LOOSE 10ML SYRINGES AND ONE PHOTO WERE RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. A VISUAL INSPECTION WAS PERFORMED. THERE WAS A LARGE RING OF SMEARED INK NOTED AROUND THE BD LOGO AND WAS NEARLY IDENTICAL ON ALL THE SYRINGES. THE OBSERVED CONDITION IS NON-CONFORMING PER PRODUCT SPECIFICATION AND IS ASSOCIATED WITH THE SCALE MARKING PROCESS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 301029, LOT 3122541. THE REVIEW REVEALED ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER APPLICABLE OPERATIONAL QUALITY SPECIFICATIONS. THERE WAS ONE QUALITY NOTIFICATION NOTED THAT PERTAINED TO THE INK RING COMPLAINT AND THE DEFECTIVE PRODUCT WAS SCRAPPED. THIS CONDITION IS OCCURRING AT OR BELOW ITS EXPECTED FREQUENCY; THEREFORE, NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. LOT 3122541 IS CONSIDERED IN COMPLIANCE WITH OUR PRODUCT SPECIFICATION REQUIREMENTS. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

POST INVESTIGATION FINDINGS REVEAL THAT THE BD SYRINGE 10ML LL BNS HAD A SCALE MARKING ISSUE. IT WAS REPORTED BY THE CUSTOMER THAT HAS BLACK MARKINGS AROUND THE SYRINGE. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED CUSTOMER REPORTED NON-CONFORMING PRODUCT FOUND IN OUR WAREHOUSE DURING PICKING PROCESS. PART # 301029, LOT # 3122541 HAS BLACK MARKINGS AROUND THE SYRINGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
377483 BD SYRINGE 10ML LL BNS PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 3122541 50382903010290

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown