FDA Adverse Event
Malfunction
Summary report: N
VASOVIEW HEMOPRO EVH SYSTEM
MDR report key: 1873691
·
Received September 21, 2010
Report
- Report Number
- 2242352-2010-02873
- Event Type
- Malfunction
- Date Received
- September 21, 2010
- Date of Event
- August 24, 2010
- Report Date
- August 24, 2010
- Manufacturer
- MAQUET CARDIOVASCULAR
- Product Code
- GEI
- PMA / PMN Number
- K052274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PHYSICIAN ASSISTANT
Narratives
Additional Manufacturer Narrative · 1
METHOD: THE DEVICE WILL REPORTEDLY NOT BE RETURNED TO MAQUET CARDIAC SURGERY FOR INVESTIGATION. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE RETURNED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE WHICH COULD BE ATTRIBUTED TO THIS FAILURE MODE. (B)(4).
Description of Event or Problem · 1
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE GRAY SILICON INSULATION (JAW BOOT) ON THE VASOVIEW HEMOPRO ENDOSCOPIC VESSEL HARVESTING SYSTEM DISLODGED FROM ITS PROPER PLACEMENT. IT NEVER FELL OFF. THE PHYSICIAN'S ASSISTANT TRIMMED THE PIECE AND COMPLETED THE CASE USING THIS DEVICE. THE PRODUCT WAS DISCARDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASOVIEW HEMOPRO EVH SYSTEM | ENDOSCOPIC VESSEL HARVESTING SYSTEM | GEI | MAQUET CARDIOVASCULAR | VH-3000 | 25014150 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |