FDA Adverse Event Injury Summary report: N

PUMP MMT-722NAS PRDGM INS V2.2 SK EN

MDR report key: 1873667 · Received October 13, 2010

Report

Report Number
3004209178-2010-83083
Event Type
Injury
Date Received
October 13, 2010
Date of Event
October 1, 2010
Report Date
October 1, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING HOSPITALIZED FOR HIGH BLOOD GLUCOSE. THE BLOOD GLUCOSE READING WAS OVER 600 MG/DL. TROUBLESHOOTING WAS PERFORMED. THE PROGRAMMING ON THE INSULIN PUMP WAS CORRECT. THE BASAL AND THE BOLUS MATCHED TO THE DAILY TOTALS. RAN A FIXED PRIME TEST AND THE INSULIN EXITED. PERFORMED A HIGH PRESSURE TEST AND THE DEVICE FAILED THE TEST. THE CUSTOMER STATED THAT SHE HAD BENT CANNULAS EVERY FEW MONTHS. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 27 YR Hospitalization