FDA Adverse Event Malfunction Summary report: N

MICRO DRILL LONG STRAIGHT ATTACHMENT

MDR report key: 1873612 · Received September 17, 2010

Report

Report Number
1811755-2010-01205
Event Type
Malfunction
Date Received
September 17, 2010
Date of Event
August 24, 2010
Report Date
August 31, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE ATTACHMENT WAS RECEIVED AT THE MANUFACTURER FOR EVALUATION, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS CONFIRMED. THE DEVICE WAS DISASSEMBLED, AND CORROSION AND DEBRIS WERE FOUND IN THE NOSE OF THE DEVICE. BASED ON THE INVESTIGATION DETAILS, THE FAILURE WAS MOST LIKELY CAUSED BY CORROSION BUILD UP IN THE ATTACHMENT DURING USE. THE FRONT OF THE ATTACHMENT HAS A GAP BETWEEN THE BUR PILOT AND THE OUTER HOUSING, WHICH HAS THE POTENTIAL TO ALLOW CORROSION AND DEBRIS TO ENTER THE ATTACHMENT. ONCE ENOUGH CORROSION BUILDS UP IN THE ATTACHMENT IT MAY CAUSE THE DEVICE TO NOT WORK PROPERLY OR CAN PHYSICALLY BIND UP THE ATTACHMENT. THE ATTACHMENT WAS SCRAPPED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATTACHMENT BEGAN OVERHEATING DURING A SURGICAL PROCEDURE. THE CASE WAS COMPLETED SUCCESSFULLY USING A BACK-UP DEVICE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO DRILL LONG STRAIGHT ATTACHMENT SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC DZJ STRYKER INSTRUMENTS KALAMAZOO 09339

Patients

Seq Age Sex Outcome Treatment
1 UNK