Report
- Report Number
- 1628664-1998-00079
- Event Type
- Other
- Date Received
- September 11, 1998
- Date of Event
- August 12, 1998
- Report Date
- September 10, 1998
- Manufacturer
- ABBOTT MANUFACTURING, INC.
- Product Code
- LCI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
ON 08/12/1998 THE ACCOUNT REPORTED A BHCG RESULT OF 2688 MIU/ML FOR A SAMPLE TESTED WITH "IMX" BHCG" REAGENT LIST NUMBER 1A06-20. THE ACCOUNT RETESTED THE SAMPLE WITH "IMX BHCG" REAGENT LIST NUMBER 1A06-22 AND OBTAINED RESULTS OF 6090/5728 MIU/ML. THE SAMPLE WAS RETESTED WITH IMX BHCG" REAGENT LIST NUMBER 1A06-20 AND A RESULT OF 2748 MIU/ML WAS OBTAINED. ALL TESTING WAS PERFORMED WITH THE AUTOMATED 1:200 DILUTION. THE CUSTOMER THEN SENT THE SAMPLE TO A REFERENCE LAB FOR TESTING AND A RESULT OF 19,089 MIU/ML WAS OBTAINED. ACCORDING TO THE ACCOUNT, AFTER SVC CHANGED THE WORN PUMP BELT ON THE INSTRUMENT, THE BHCG RESULTS CORRELATED BETWEEN BOTH "IMX" REAGENT LIST NUMBERS AND THE REFERENCE LAB METHODOLOGY. THE EXACT CONCENTRATIONS OBTAINED WERE NOT PROVIDED BY THE ACCOUNT. NO SAMPLE WAS AVAILABLE FROM THE REFERENCE LAB TO VERIFY TEST RESULTS AFTER MAINTENANCE PERFORMED. NO REPORT OF INJURY DUE TO THE INITIAL REPORTED RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMX ANALYZER | AUTOMATED IMMUNOASSAY ANALYZER | LCI | ABBOTT MANUFACTURING, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR | IMXB HCG REAGENTS, LIST NUMBERS| 1A06-20 AND 1A06-22. |