FDA Adverse Event
Malfunction
Summary report: N
U2 DRILL
MDR report key: 1873502
·
Received October 14, 2010
Report
- Report Number
- 1811755-2010-01395
- Event Type
- Malfunction
- Date Received
- October 14, 2010
- Date of Event
- September 17, 2010
- Report Date
- September 22, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DRILL WAS RECEIVED BY THE MFR AND THE COMPLAINT WAS CONFIRMED. INTERNAL COMPONENTS WERE EVALUATED, WHERE EXTENSIVE CORROSION WAS DISCOVERED ON THE ROTOR AND BEARINGS. THE PRESENCE OF CORROSION CAN LEAD TO INCREASED FRICTION BETWEEN ROTATING COMPONENTS, RESULTING IN HEAT BEING GENERATED. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A DEVICE EVALUATION CONDUCTED AT THE MFR, THE DRILL HEATED UP. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | U2 DRILL | DRILLS, SURGICAL, ENT (ELECTRIC OR PNEUMATIC), INC | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |