FDA Adverse Event Malfunction Summary report: N

U2 DRILL

MDR report key: 1873502 · Received October 14, 2010

Report

Report Number
1811755-2010-01395
Event Type
Malfunction
Date Received
October 14, 2010
Date of Event
September 17, 2010
Report Date
September 22, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DRILL WAS RECEIVED BY THE MFR AND THE COMPLAINT WAS CONFIRMED. INTERNAL COMPONENTS WERE EVALUATED, WHERE EXTENSIVE CORROSION WAS DISCOVERED ON THE ROTOR AND BEARINGS. THE PRESENCE OF CORROSION CAN LEAD TO INCREASED FRICTION BETWEEN ROTATING COMPONENTS, RESULTING IN HEAT BEING GENERATED. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DEVICE EVALUATION CONDUCTED AT THE MFR, THE DRILL HEATED UP. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 U2 DRILL DRILLS, SURGICAL, ENT (ELECTRIC OR PNEUMATIC), INC ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK