FDA Adverse Event Malfunction Summary report: N

REPAIR U2 DRILL

MDR report key: 1873481 · Received October 14, 2010

Report

Report Number
1811755-2010-01400
Event Type
Malfunction
Date Received
October 14, 2010
Date of Event
September 23, 2010
Report Date
September 23, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DRILL WAS RECEIVED BY THE MANUFACTURER, HOWEVER, THE OVERHEATING COMPLAINT COULD NOT BE REPLICATED. DURING THE DEVICE EVALUATION, IT WAS NOTICED THAT THE DEVICE WAS EXCESSIVELY NOISY, PROMPTING FURTHER INVESTIGATION. INTERNAL COMPONENTS WERE EVALUATED AND THE ROTOR BEARINGS WERE DISCOVERED TO BE MISSING THE REQUIRED LUBRICATION. IF LUBRICATION IS NOT PRESENT, HEAT COULD BE GENERATED AMONG ROTATING COMPONENTS. THE ROTOR ASSEMBLY AND BEARINGS WERE REPLACED, AND ADDITIONAL PREVENTATIVE MAINTENANCE WAS ALSO PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED PRIOR TO THE START OF A SURGICAL PROCEDURE, THE DRILL HEATED UP. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO DELAY TO THE SCHEDULED PROCEDURE. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REPAIR U2 DRILL DRILLS, SURGICAL, ENT (ELECTRIC OR PNEUMATIC), INC ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK