FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1873466 · Received September 20, 2010

Report

Report Number
1720753-2010-03133
Event Type
Malfunction
Date Received
September 20, 2010
Date of Event
September 3, 2010
Report Date
September 20, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYS AND FOUND A BROKEN WIRE IN THE INTERCONNECT CABLE. THE CABLE WAS ORDERED AND THE CUSTOMER INSTALLED THE NEW CABLE. SYS OPERATES AS INTENDED.

Description of Event or Problem · 1

CUSTOMER REPORTED THE SYS IS DISPLAYING A SYSTEM ERROR "POWER OFF WAIT 10 SEC" ON THE MAINFRAME DISPLAY AND INTERMITTENTLY THE C-ARM WOULD NOT BOOT. IT WOULD STOP AT 5 ARROWS ON THE DISPLAY. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1