FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3000 SYSTEM

MDR report key: 1873447 · Received October 14, 2010

Report

Report Number
3002158293-2010-01065
Event Type
Malfunction
Date Received
October 14, 2010
Date of Event
September 29, 2010
Report Date
October 13, 2010
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. UPON EVALUATION THE CONNECTOR ON THE TRUNK CABLE WAS BROKEN. THE ROOT CAUSE OF THE BROKEN CONNECTOR CANNOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY A RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT.

Description of Event or Problem · 1

A REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING ELECTRODE BELT (B)(4) WAS FOUND TO HAVE A BROKEN CONNECTOR ON THE TRUNK CABLE. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 3000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 3000 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK